PFG-2085 Industrial Online Fluorine Ion Tester
PFG2085 Industrial online fluorine ion analyzer is our companys latest, intelligent fluorine ion analyzer. It is designed to
measure fluorine ion online monitoring. It is suitable for the measured in numerical fluoride occasion of containing fluoride
ion.
Features
English LCD display, English menu operation, English notepad: It adopt menu structure similar to computer |
Operation, simple operation, the steps throughout the English tips, without instructions, it can be completed |
Multi-parameter display: The concentration value, temperature, time and status can be displayed on the screen |
at the same time. |
Intelligent: On-line Fluorine Ion Detector adopts high-precision AD conversion and single chip microcomputer |
processing technologies. It can be used for the measurement of the Fluorine ion concentration value and |
temperature, automatic temperature compensation, self-checking etc various functions |
Reliability: All the components are arranged on one circuit board. No complicated functional switch and |
adjusting knob |
Strong interference immunity: Its current output adopts optoelectronic isolating technology. This meter has |
strong interference immunity and the capacity of long-distance transmission as well as excellent |
electromagnetic compatibility |
Automatic timing calibration: It can ensure the stability and reliability of the measurement instrument |
Instrument not crash: Watchdog procedures to ensure that the instrument will not working continuously crash |
Technical Indexes
1. Measuring range: 0.2ug/L~2300mg/L temperature:0~99.9℃ |
2. Resolution: 0.1ug/L, temperature:0.1℃ |
3. Automatic temperature compensation range:: 0~99.9℃,with 25℃ as the reference temperature |
4. Accuracy:±2.5%,±0.3℃ |
5. Water sample tested:: 0~99.9℃, 0.6MPa |
6.Stability:±1.5%/24h |
7.Response speed:32/s 25℃,50/s 15℃ |
8. Clock accuracy: ±1 minute/month |
9. Isolated current output: 0-10mA (load resistance <15KΩ); |
4-20mA (load resistance <750Ω) |
10. Output current error: ≤±1%FS |
11. Data storage capacity: l month (1 point/5 minutes) |
12. High and low alarm relay: AC220V, 3A |
13. Communication interface: RS485 |
14. Power supply: AC220V±22V, 50Hz±1Hz |
15. Protection grade: lP65 |
16. Overall dimension: 146 (length) x 146 (width) x 108 (depth) mm; |
dimension of the hole: 138 x 138mm |
17. Weight: 0.8kg |
18. Working conditions: ambient temperature: 0-60℃ |
relative humidity: <85% |
Attention:
Output signal 4~20mA: current output of measured fluorine ion valueIf you need a temperature corresponding to
the 4~20mA current output of dual output, because of requires special order, you must specify when ordering.