Product Description
The instrument can be used for the measurement of dissolved oxygen concentration,
saturation, electrode current and temperature. It adopts the microprocessor
technology and wide-screen LCD display. At the same time, it also owns
the operation prompts, making operation simple and convenient.
Features
♦ It is available with automatic temperature compensation. It is also
featured by zero oxygen, full-scale, air pressure, power-off protection
and other functions.
♦ It is with the power-off protection function.
♦ The instrument adopts the low-power consumption design, with
low-voltage display, automatic shutdown and other power
management functions.
♦ The protection grade of the instrument casing is lP65. It is waterproof,
dustproof and suitable for field work.
♦ It is equipped with the dissolved oxygen electrode (polarographic
membrane oxygen) made by our factory.
Technical Indexes
1. Measuring range: dissolved oxygen: 0.00-19.99mg/L
Dissolved oxygen saturation: 0.0 - 199.9 %
Temperature: 0 .0 - 60.0 °C
2. Automatic temperature compensation: 0~60°C
3. Intrinsic error of the instrument:
Dissolved oxygen: ± 0.1 ug/L ± 1 unit (at the calibration temperature)
± 0.5ug/L (when the difference is ± l 5 °C, compared with the calibration
temperature)
Dissolved oxygen saturation: ± 0 5 0% (at the calibration temperature)
± 1.0% (when the difference is ± l 5 °C, compared with the calibration
temperature)
Temperature: ± 0 .5 °C ± l unit
4. Residual current: 0.1 ug/L
5. Response time: 20s (at 20 °C, 90% response)
6. Instrument stability: ± 0.2ug/L/1h
7. Instrument repeatability: 0.2ug / L
8. Instrument weight: 0.3kg
9. Overall dimension: 210 ×100 ×45mm
10. Normal operating conditions of the instrument
a) Ambient temperature: (0 - 40) °C
b) Environmental adaptability: In line with the shell protection grade IP65
c) Measured solution temperature: (0 - 40) °C
d) Power supply: One 9V battery
e) Except for the geomagnetic field around, there is no other vibration
affecting the performance.