DOG-2082 Industrial online Dissolved Oxygen Meter
DOG-2082 Industrial online Dissolved Oxygen Meter is a kind of high-intelligence on line meter that is developed by our company and used for continuous on line monitoring. It can be equipped with DOG-208F Polarographic Electrode and can automatically switch from ppb level to ppm level of wide-range measurement. This instrument is designed for monitoring the oxygen content in the boiler feed water, condensate water and sewage.
Features
English menu operation: |
It has a complete English display and elegant interface: Liquid crystal display module with high resolution is |
adopted. All the data, status and operation prompts are displayed in English. |
Multi-parameter display: The oxygen concentration value, input current (or output current), temperature values, |
time and status can be displayed on the screen at the same time. concentration value in a 10 x 10mm size. As the |
main display is eye-catching, the displayed values can be seen from a long distance away. Six sub-displays can |
display such information as the input or output current, temperature, status, week, year, day, hour, minute and |
second, so as to adapt to different users’ habits and to conform with different reference times set by the users. |
Technical Indexes
1. Measuring range: 0~100.0ug/L; 0~20.00 mg/L (automatic switching); 0~60% |
Resolution: 0.1ug/L; 0.01 mg/L; 0.1°C |
2. Intrinsic error of the whole instrument: ug/L: ±l.0%FS; mg/L: ±0.5%FS, temperature: ±0.5°C |
3. Repeatability of the indication of the whole instrument: ±0.5%FS |
4. Stability of the indication of the whole instrument: ±1.0%FS |
5. Automatic temperature compensation range: 0~60°C, with 25°C as the reference temperature. |
6. Response time: <60s (98% and 25°C of the final value) 37°C: 98% of the final value < 20 s |
7. Clock accuracy: ±1 minute/month |
8. Output current error: ≤±l.0%FS |
9. Isolated output: 0-10mA (load resistance <15KΩ); 4-20mA (load resistance <750Ω) |
10. Communication interface: RS485 or 232(optional) |
11. Data storage capacity: l month (1 point/5 minutes) |
12. Saving time of data under a continuous power-failure condition: 10 years |
13. Alarm relay: AC 220V, 3A |
14. Power supply: AC220V±10%, 50±1HZ, 24V DC (optional) |
15. Overall dimension: secondary meter: 146 (length) x 146 (width) x 108 (depth) mm; |
dimension of the hole: 138 x 138mm |
16. Weight: 0.8.kg ( meter) |
17. Working conditions: ambient temperature: 0-60°C; relative humidity <85% |
18. The connection tubes for inlet and outlet water: Pipes and hoses should be of the following three sizes: Φ8, |
Φ10 and Φl2. |
|