XY-1HWT Type High-frequency Infrared Carbon & Sulfur Analyzer compatible applying with Type High frequency automatic inductive combustion furnace, can fleetly and exactly measure the carbon and sulfur in steel, iron, alloy, nonferrous metals, cement, ore and other materials. This equipment is a high-tech products integrate light, motor, electronic, computer and analysis technology, with features of wide measuring scope, strong anti interference, multiple function, easy to operate, accurate and reliable analysis result.
Main Parameter:
Measure Range:
Carbon: w(C) 0.0001%-10.0000% (can be extended to 99.999%)
Sulpher: w(C) 0.0001%-3.5000% (can be extended to 99.999%)
Sample weight:
standard 0.5g
Analysis Error:
Carbon: accord with ISO9556 standard Sulfur: accord with ISO4935 standard
Analysis Precision: Carbon: RSD<0.5% Sulfur: RSD<1%
Analysis pool
Carbon: low carbon pool and high carbon pool
Sulfur: if the customer request,the high sulfur pool could be added
Time of analysis:
25 to 60 seconds, could be adjusted. Around 35 seconds usually
Sensitivity(The minimum readings):
C/s 0.1ppm
Electronic scale:
Precision of read: 0.0001g
Working Environment:
Room temp: 10 to 30 degree C,relative Humidity less than 75%
Configuration of the instrument
Infrared Detection Unit
Circuit Design: The modularized design with double CPU for the whole frame is divided into the upper and lower levels while the electronic circuits are highly integrated stably and reliably; in the meantime the multilevel hidden-type isolated circuit is employed so that it is free from HF interference.
Power: The linear close-coupled modular power block ensures stable output without fault.
Light source: The tailor-made new model platinum infrared light source has lasting caloricity and efficient spectral property.
Analysis cell: The gilded carbon & sulfur analysis cell is fit with the high precision TSC infrared detector (10-11); it is also possible to have the double-carbon cell or the double- carbon & double-sulfur cell fit out at the client’s request.
Motor: The special-purpose aviation modulation motor is applied with satisfactory thermal stability whose life expectancy exceeds 100,000 hours.
High-frequency combustion system
Radio circuit: The design of high-duty radio circuit and the application of 2.5KVA HF pliotron. frequency : 20MHz
Military-purpose ceramic vacuum tubes and ceramic vacuum capacitors.
HF control circuit: It is used for automatic detection of the electromagnetic valve, the elevation or descent of cylinder and the performance of HF unit.
Automatic overtime/overflow alarming system enables the HF furnace to work under normal condition.
The optional current/voltage/power regulator for furnace temperature control: It is applicable to samples of various materials.
Analytical gas path system
Gas path: The high precision flow controller ensures the stability of gas flow as well the gas intake system (such as electromagnetic valves, unions, cylinder hoists) for automatic leakage detection.
Dual standard correction for solid and gas
De-dusting unit: The combustion head self-cleaning device effective for reducing the influence of dust for the result of analysis; ash removing system for the inlet.
0.4 μm submicron metal filter secures thorough separation of dust from gas and can be used for a long time with no need of the ultrasonic cleaner.
Input/Output System
Computer
Printer: HP
Electronic 110g/0.1mg
operational software
Operational software: Adopt Deliph software compiler , WINDOWS XP english operating
software.
Analysis channel: Providing channel management features, the carbon and sulfur channels are free to increase, delete and edit, no limit. Preset the interface of low-carbon and sulfur, Can realize the software automatically change
Analysis function: Analysis of the dynamic data, sampling every 20 times, improving the analysis sensitivity and analysis accuracy, providing sample management features, it can edit the sample name and logo, also can increase and delete samples, the software provides a user management system, the administrator setting different user permissions.
Display function: The Carbon sulfur has a curve respectively,
Data processing function: Adopt ACCESS data pool store the analysis results, can store all data and curve;
Query the analysis results optionally, query according to the time, sample name, logo and operator.
Provide some function as data storage, blank deduction, setting parameter, choosing channel, statistics, and curve Compare etc.
In the software can create the work curve of carbon and sulfur, realizing the curve fitting.
Printing function: Print mode is diversification, Provide two print modes of laboratory and testing, also can design printing formats
self-diagnose function: System diagnosis function,can test furnace head and the gas chamber sealing by the software.
Standard Accessories
1.Carbon/Sulfur analyzer
Item Description Q’ty
1 Main unit XY-1HW 1
2
High frequency automatic inductive combustion furnace
1
3 Electric Balance
110g / 0.1mg
1
4 synopsis 1
2. Spare parts list
DEscriptION SPECIFICATION QUANTITY
Relieve Valve 1 Piece
Tungsten fluxing medium 1 Kilogram
Pure Iron fluxing medium 2 Bottle
Quartz Combustion Tube 2 Piece
De-dusting Tube 2 Piece
Copper Crucible Holder 1 Piece
Ceramic Crucible Holder 2 Piece
Ceramic Crucible 1000 piece /box 1 Box
Combustion Tube Brush Cleaner 5 Piece
Steel Wire Brush 2 Piece
Forceps 1 Pair
Sampling Spoon 1 Set
Tools 1 Set
PU Pipe 4mm×6mm 10 M
Silicon Ring 41mm×6.5mm 2 Piece
O-Ring 11mm×2.2mm 4 Piece
O-Ring 36mm×3mm 4 Piece
Vacuum Silicone Grease 1 Bottle
Fuse 4A,20A 4 for each
Quick Plug PC06-01 2 Piece
Quick Plug PC06-02 2 Piece
Multi-socket 220V / 10A 1 Piece
Desiccant 125g/bottle 1 Bottle
CO2 Absorbent 125g/bottle 1 Bottle
discharging dust tube Silicon Rubber 2 Piece
Packing & Transportation
Wooden packing in conformity to GB15464-95
shipping by sea or air transport
Other items
Overall repair warranty for 2 years, together with lifelong maintenance, free installation and commission and free training the lab technicians on site as well as providing the analysis methods.
Acceptance terms
Preparations
When the instruments arriving, user should prepare install place, power (earthing cable) oxygen, dynamic gas and indoor environment
With the above conditions, please written notice to Party B of the installation engineer to the site installation and commissioning.
Open-package inspection instruments
1.The instruments should with packing list, the instrument should is new and no breakage.
2.Confirm the all equipment accessories according to the scope of supply.
3.The installation engineer of party A should in when open the package, if the equipment is shortages, the engineer should confirm and complete in 30 days.
acceptance standard
1.Repeatability
Select steel of one standard sample (Carbon mass fraction in 0.400%-1.000%,sulfur content in 0.020% - 0.080%) operating according to instrument instructions, with high-frequency furnace
continuous testing 7 times, The difference between the maximum and minimum of test results should be less than or equal to 1.5r. The r with the following formula to calculate:
Carbon:lgr = 0.52121lgω-1.9617
sulfur:lgr = 0.62212lgω-1.9965
the r mean repeatability, ω mean the value of the material under test.
2.Accuracy
Select steel or iron of standard sample(carbon mass fraction in 0.003% - 4.500%, sulfur mass fraction in 0.002% - 0.100%) take the high, medium and low three kinds of samples, operating
according to the instruction of high-frequency furnace and correction, after correction every standard material test 3 times, taking the average, carbon and sulfur should be less than or equal
to the calculated values of
CrD95 X − X0
,see the following formula:
X mean the average
Xo mean the value of the material under test
n mean test times
r mean repeatability
R mean reproducibility
The reproducibility with the following formula to calculate: carbon:lgR = 0.58221lgω-1.4558
sulfur: lgR = 0.68878lgω-1.3693
ω mean the value of the material under test.
Training
Scene technical training
Party A is responsible for party B of the operation and maintenance, daily breakdown training, ensure the worker of party B can operate instrument independently.
System training
The party B can send staff to party A system training, two staffs, five days(Training specific time agreed both sides) transport, accommodation self-care. Training includes:
Principles and methods of instrumental analysis
The structure and features of the instrument
Quantitative analysis of instrument (iron and steel material)
Analysis conditions set, the working curve, correction, etc.
Daily breakdown exclusion
Instrument Maintenance
Installation of the instrument Preparation before the installation
1.Analyzing room: be far away from the erosion gas such as acid, alkali, dust, vibration, testing disturbance.
2.Analyzing room size: >3×3m;
3.Working condition: Room temperature: 10-30℃, Relative humidity: <75%;
4.Power supply: well earthing is required, Voltage: AC220V±5%, Frequency: 50Hz±2%, No disturbance from harmonic wave
5.Voltage stabilizing equipment: Power: 5kW, Voltage stabilizing precision<2%;
Note: If there is disturbance from intermediate frequency furnace in the work power supply, the equipment should be equipped with AC Regulated Power Supply.
6.Gas: Oxygen: purity≥99.5%
7.Power gas: Nitrogen or Compressed air (Purified: without water, oil and contamination)
8.Tool: Box opening tool and regular tools.